Investigation of the HotDog patient warming system: detection of thermal gradients
Document Type
Article
Publication Date
5-1-2018
Abstract
Objective: To assess the performance of an active patient-warming device. Materials and Methods: Temperatures of an active patient-warming device (HotDog system) were measured at various time points using an infrared thermometer. The study was conducted in two phases: Phase 1 compared temperatures among four different areas of the warming blanket. Phase 2 compared conditions simulating different scenarios using a weighted patient simulator. Results: Phase 1: Three out of four positions on the warming blanket had significantly different temperature measurements. Phase 2: Temperature output by the warming blanket was reduced: (1) in the absence of the patient simulator placed across the blanket (−1·9°C, P=0·013); (2) if the patient simulator was placed away from the blanket sensor (−2·0°C, P=0·009); and (3) if there was fluid between the patient simulator and warming blanket (−2·2°C, P=0·004). In a majority of measurements (95%), the set temperature of 43°C on the control unit was not reached (range, 29·8 to 42·9°C) and 2·3% of measurements were higher (range, 43·1 to 45·8°C) than the control unit set temperature of 43°C. Clinical Significance: Measured temperatures on the active warming blanket did not reflect control unit settings. This could result in the potential for hyperthermic injury, ineffectual heating and uneven heat distribution.
Publication Title
Journal of Small Animal Practice
Volume
59
Issue
5
First Page
298
Last Page
304
Digital Object Identifier (DOI)
10.1111/jsap.12816
PubMed ID
29363138
ISSN
00224510
E-ISSN
17485827
Citation Information
McCarthy, Matz, B., Wright, J., & Moore, L. (2018). Investigation of the HotDog patient warming system: detection of thermal gradients. Journal of Small Animal Practice, 59(5), 298–304. https://doi.org/10.1111/jsap.12816